A bootstrap method for gain calibration and resolution determination of a lead-glass calorimeter
Oct 15, 2006
9 pages
Published in:
- Nucl.Instrum.Meth.A 566 (2006) 366-374
Report number:
- JLAB-PHY-06-527
View in:
Citations per year
Abstract: (Elsevier)
We describe a method for calibration of a lead-glass calorimeter that does not require a beam of known energy. The technique was used to calibrate the radphi lead-glass calorimeter at Jefferson Lab. The technique described can be applied to any segmented electromagnetic calorimeter capable of detecting all-photon decays of mesons, for example, π 0 → 2 γ , η → 2 γ or ω → π 0 γ . We also demonstrate how the measured 2 γ mass width of the π 0 and η mesons can be unfolded to extract the single-shower energy and position resolution functions of the calorimeter.- 14.40.Cs
- 25.20.Lj
- Meson
- Photoproduction
- Radiative
- Decay
- Phi
- Radphi
- Calorimeter
- Calibration