Upgraded photon calorimeter with integrating readout for Hall A Compton Polarimeter at Jefferson Lab
Aug, 2011
16 pages
Published in:
- Nucl.Instrum.Meth.A 676 (2012) 96-105
e-Print:
- 1108.3116 [physics.ins-det]
Report number:
- JLAB-PHY-11-1357
Citations per year
Abstract: (arXiv)
The photon arm of the Compton polarimeter in Hall A of Jefferson Lab has been upgraded to allow for electron beam polarization measurements with better than 1% accuracy. The data acquisition system (DAQ) now includes an integrating mode, which eliminates several systematic uncertainties inherent in the original counting-DAQ setup. The photon calorimeter has been replaced with a Ce-doped GSO crystal, which has a bright output and fast response, and works well for measurements using the new integrating method at electron beam energies from 1 to 6 GeV.Note:
- 16 pages, 15 figures
- Compton polarimeter
- Jefferson Laboratory
- Electron beam polarimetry
- Integrating DAQ
- photon: calorimeter
- electron: polarized beam
- data acquisition
- Jefferson Lab
- polarization
- Compton scattering: backscatter
References(17)