Design, Construction, Operation and Performance of a Hadron Blind Detector for the PHENIX Experiment
Mar, 2011
Citations per year
Abstract: (arXiv)
A Hadron Blind Detector (HBD) has been developed, constructed and successfully operated within the PHENIX detector at RHIC. The HBD is a Cherenkov detector operated with pure CF4. It has a 50 cm long radiator directly coupled in a window- less configuration to a readout element consisting of a triple GEM stack, with a CsI photocathode evaporated on the top surface of the top GEM and pad readout at the bottom of the stack. This paper gives a comprehensive account of the construction, operation and in-beam performance of the detector.Note:
- 51 pages, 39 Figures, submitted to Nuclear Instruments and Methods
- 29.40.-n
- 29.40.Cs
- 29.40.Ka
- 25.75.-q
- HBD
- GEM
- CsI photocathode
- UV-photon detector
- CF 4
- detector: performance
References(24)