Comparison of commercial thermoluminescent readers regarding high-dose high-temperature measurements

Jan 24, 2014
6 pages
Published in:
  • Radiat.Meas. 65 (2014) 8-13
  • Published: 2014
e-Print:

Citations per year

20112012201301
Abstract: (arXiv)
Three different thermoluminescent measuring systems have been compared with respect to the differences in temperature profiles, spectral sensitivities, as well as linearity of dose response characteristics. The comparison was performed using the Harshaw 3500, Riso DA-20 and RA94 TLD readers. The instruments were tested for the readouts of highly irradiated LiF:Mg,Cu,P (MCP) TL detectors, which require readout up to 550 C, in case of doses exceeding 1 kGy. It was found that the Harshaw 3500 can be used, without any additional light attenuation, for the measurements of MCP detectors exposed to doses up to about 5 Gy. For the other two readers the upper dose limit is about 5 times lower. It was also found that the Harshaw 3500 shows the best thermal stability considering the peak maximum position. For the ultra-high doses the differences in the spectral characteristics of the applied optical filters and photomultipliers, in conjunction with an evolution of the MCP TL emission spectrum with increasing dose, significantly influence the shape of TL glow curves measured with the Riso reader. The detailed characteristic of the compared TLD readers, at high-dose high-temperature range is discussed.
Note:
  • Data contained within this manuscript have been presented during the 1st International Conference on Dosimetry and its Applications, Prague, Czech Republic, 23-28 June 2013
  • LiF:Mg,Cu,P
  • High temperatures
  • High doses
  • TLD readers
  • Comparison
  • [Bilski et al, 2008]

    Characteristic of LiF: Mg,Cu, P thermoluminescence at ultra-high doses range

    • P. Bilski
      ,
    • B. Obryk
      ,
    • P. Olko
      ,
    • E. Mandowska
      ,
    • A. Mandowski
    et al.
      • Radiat.Meas. 43 (2008) 315
  • [Bøtter-Jensen et al, 2000]

    Advances in luminescence instrument systems

    • L. Bøtter-Jensen
      ,
    • E. Bulur
      ,
    • G.A.T. Duller
      ,
    • A.S. Murray
      • Radiat.Meas. 32 (2000) 523
  • [Gieszczyk et al, 2013a]

    Spectral characteristic of high-dose high temperature emission from LiF: Mg,Cu, P (MCP-N) TL detectors

    • W. Gieszczyk
      ,
    • P. Bilski
      ,
    • B. Obryk
      ,
    • P. Olko
      ,
    • A.J.J. Bos
      • Radiat.Meas. 53-54 (2013) 22
  • [Gieszczyk et al, 2013b]

    Measurements of high-temperature emission spectra of highly irradiated LiF: Mg,Cu, P (MCP-N) TL detectors

    • W. Gieszczyk
      ,
    • P. Bilski
      ,
    • B. Obryk
      ,
    • P. Olko
      ,
    • A.J.J. Bos
      • Radiat.Meas. 56 (2013) 183
  • [Mandowska et al, 2010]

    Spectrally resolved thermoluminescence of highly irradiated LiF: Mg, Cu, P detectors

    • E. Mandowska
      ,
    • P. Bilski
      ,
    • B. Obryk
      ,
    • A. Mandowski
      ,
    • P. Olko
    et al.
      • Radiat.Meas. 45 (2010) 579
  • [Obryk et al, 2009]

    Development of a method for passive measurement of radiation doses at ultra-high dose range

    • B. Obryk
      ,
    • P. Bilski
      ,
    • M. Budzanowski
      ,
    • M. Fuerstner
      ,
    • M. Glaser
    et al.
      • IEEE Trans.Nucl.Sci. 56 (2009) 3759
  • [Obryk et al, 2011a]

    Method of thermoluminescent measurement of radiation doses from micrograys up to a megagray with a single LiF: Mg,Cu, P detector

    • B. Obryk
      ,
    • P. Bilski
      ,
    • P. Olko
      • Radiat.Prot.Dosimetry 144 (2011) 543
  • [Obryk et al, 2011b]

    Response of various types of lithium fluoride MCP detectors to high and ultra-high thermal neutron doses

    • B. Obryk
      ,
    • M. Glaser
      ,
    • I. Mandić
      ,
    • P. Bilski
      ,
    • P. Olko
    et al.
      • Radiat.Meas. 46 (2011) 1882
  • [Obryk, 2013]

    From nGy to MGy – new dosimetry with LiF: Mg,Cu, P thermoluminescence detectors

    • B. Obryk