Surface and buildup region dose measurements with Markus parallel-plate ionization chamber, Gafchromic EBT3 film and MOSFET detector for high-energy photon beams

Jan 14, 2016
10 pages
Published in:
  • Adv.High Energy Phys. 2016 (2016) 8361028
  • Published: 2016
e-Print:

Citations per year

0 Citations
Abstract: (HINDAWI)
The aim of the study was to investigate surface and buildup region doses for 6 MV and 15 MV photon beams using a Markus parallel-plate ionization chamber, GafChromic EBT3 film, and MOSFET detector for different field sizes and beam angles. The measurements were made in a water equivalent solid phantom at the surface and in the buildup region of the 6 MV and 15 MV photon beams at 100 cm source-detector distance for 5 × 5, 10 × 10, and 20 × 20 cm^2 field sizes and 0, 30, 60, and 80 beam angles. The surface doses using 6 MV photon beams for 10 × 10 cm^2 field size were found to be 20.3%, 18.8%, and 25.5% for Markus chamber, EBT3 film, and MOSFET detector, respectively. The surface doses using 15 MV photon beams for 10 × 10 cm^2 field size were found to be 14.9%, 13.4%, and 16.4% for Markus chamber, EBT3 film, and MOSFET detector, respectively. The surface dose increased with field size for all dosimeters. As the angle of the incident radiation beam became more oblique, the surface dose increased. The effective measurement depths of dosimeters vary; thus, the results of the measurements could be different. This issue can lead to mistakes at surface and buildup dosimetry and must be taken into account.
  • [1]

    Accurate skin dose measurements using radiochromic film in clinical applications

    • S. Devic
      ,
    • J. Seuntjens
      ,
    • W. Abdel-Rahman
      • Med.Phys. 33 (2006) 1116-1124
  • [2]

    Entrance and exit dose regions for a clinac-2100c

    • E.E. Klein
      ,
    • J.A. Purdy
      • Int.J.Radiat.Oncol.Biol.Phys. 27 (1993) 429-435
  • [3]

    Determination of build-up region overresponse corrections for a Markus-type chamber

    • D.E. Mellenberg, Jr.
      • Med.Phys. 17 (1990) 1041-1044
  • [4]

    Radiochromic plastic films for accurate measurement of radiation absorbed dose and dose distributions

    • W.L. McLaughlin
      ,
    • A. Miller
      ,
    • S. Fidan
      ,
    • K. Pejtersen
      ,
    • W. Batsberg Pedersen
      • Radiat.Phys.Chem. 10 (1977) 119-127
  • [5]

    Radiochromic film for medical radiation dosimetry

    • M.J. Butson
      ,
    • P.K.N. Yu
      ,
    • T. Cheung
      ,
    • P. Metcalfe
  • [6]

    Surface dose measurements with GafChromic EBT film for 6 and 18 MV photon beams

    • H. Bilge
      • Phys.Medica 25 (2009) 101-104
  • [7]

    Surface dose and build-up region measurements with wedge filters for 6 and 18 MV photon beams

    • H. Bilge
      ,
    • N. Ozbek
      ,
    • M. Okutan
      ,
    • A. Cakir
      ,
    • H. Acar
      • Radiology 28 (2010) 110-116
  • [8]

    A new radiotherapy surface dose detector: the MOSFET

    • M.J. Butson
      ,
    • A. Rozenfeld
      ,
    • J.N. Mathur
      ,
    • M. Carolan
      ,
    • T.P.Y. Wong
    et al.
      • Med.Phys. 23 (1996) 655-658
  • [9]

    Measurement of radiotherapy x-ray skin dose on a chest wall phantom

    • K.Y. Quach
      ,
    • J. Morales
      ,
    • M.J. Butson
      ,
    • A.B. Rosenfeld
      ,
    • P.E. Metcalfe
      • Med.Phys. 27 (2000) 1676-1680
  • [10]

    Water equivalent properties of materials commonly used in proton dosimetry

    • P. de Vera
      ,
    • I. Abril
      ,
    • R. Garcia-Molina
      • Appl.Radiat.Isot. 83 (2014) 122-127
  • [11]

    How water equivalent are water-equivalent solid materials for output calibration of photon and electron beams?

    • V.M. Tello
      ,
    • R.C. Tailor
      ,
    • W.F. Hanson
      • Med.Phys. 22 (1995) 1177-1189
  • [12]

    Measurement of dose in the buildup region using fixed-separation plane-parallel ionization chambers

    • B.J. Gerbi
      ,
    • F.M. Khan
      • Med.Phys. 17 (1990) 17-26
  • [13]

    Central axis depth dose data for use in radiotherapy

    • E.G.A. Aird
      ,
    • J.E. Burns
      ,
    • M.J. Day
  • [14]

    Accurate dosimetry with GafChromic6 EBT film of a 6 MV photon beam in water: what level is achievable?

    • L.J. Van Battum
      ,
    • D. Hoffmans
      ,
    • H. Piersma
      ,
    • S. Heukelom
      • Med.Phys. 35 (2008) 704-716
  • [15]

    Calibration of a MOSFET detection system for 6-MV in vivo dosimetry

    • P. Scalchi
      ,
    • P. Francescon
      • Int.J.Radiat.Oncol.Biol.Phys. 40 (1998) 987-993
  • [16]

    Determination of dose equivalents resulting from external radiation sources

  • [17]

    Build up region of megavoltage photon radiation sources

    • D.E. Velkley
      ,
    • D.J. Manson
      ,
    • J.A. Purdy
      ,
    • G.D. Oliver, Jr.
      • Med.Phys. 2 (1975) 14-19
  • [18]

    Skin dose estimation for various beam modifiers and source-to-surface distances for 6MV photons

    • G. Yadav
      ,
    • R.S. Yadav
      ,
    • A. Kumar
      • Med.Phys. 34 (2009) 87-92
  • [19]

    Characterization of MOSkin detector for in vivo skin dose measurement during megavoltage radiotherapy

    • W.L. Jong
      ,
    • J.H.D. Wong
      ,
    • N.M. Ung
      • Med.Phys. 15 (2014) 120-132
  • [20]

    In vivo verification of superficial dose for head and neck treatments using intensitymodulated techniques

    • Z.-Y. Qi
      ,
    • X.-W. Deng
      ,
    • S.-M. Huang
      • Med.Phys. 36 (2009) 59-70
  • [21]

    Extrapolated surface dose measurements with radiochromic film

    • M.J. Butson
      ,
    • P.K.N. Yu
      ,
    • P.E. Metcalfe
      • Med.Phys. 26 (1999) 485-488
  • [22]
    Methods for In Vivo Dosimetry in External Radiotherapy, ESTRO Booklet no. 1, Garant, Leuven, Belgium
    • J. Van Dam
      ,
    • G. Marinello
  • [23]

    The effects of thermoplastic masks used for immobilization of patients on surface and buildup for megavoltage radiation

    • H. Elvan Erkan
      ,
    • G. Kemikler
  • [24]
    The Physics of Radiation Therapy, Lippincott Williams & Wilkins, Philadelphia, Pa, USA, 4th edition
    • F.M. Khan
  • [25]

    Skin dose measurement by using ultra-thin TLDs

    • J.-P. Lin
      ,
    • T.-C. Chu
      ,
    • S.-Y. Lin
      ,
    • M.-T. Liu
      • Appl.Radiat.Isot. 55 (2001) 383-391