Building of a 4-channel TTL scaler for counting detector signals

Jul 30, 2016
8 pages
e-Print:
Report number:
  • RD51-NOTE-2016-003
Experiments:

Citations per year

20162018202020222022102
Abstract: (arXiv)
A scaler has been fabricated to count the signals from any radiation de- tector. It can count signals of frequency up to 140 kHz. Transistor Transistor Logic (TTL) is used in this scaler. In this article the details of the design, fabrication and operation processes of the scaler is presented.
Note:
  • 8 pages, 9 figures, RD51-NOTE-2016-003
  • fabrication
  • radiation detector
  • electronics: readout
  • electronics: design
  • signal processing
  • performance