The PRad Windowless Gas Flow Target
Mar 1, 2021
Published in:
- Nucl.Instrum.Meth.A 1003 (2021) 165300
- Published: Jul 1, 2021
e-Print:
- 2103.01749 [physics.ins-det]
Report number:
- JLAB-PHY-21-3330
View in:
Citations per year
Abstract: (Elsevier)
We report on a windowless, high-density, gas-flow target at Jefferson Lab that was used to measure rp , the root-mean-square charge radius of the proton. The target achieved its design goal of an areal density of 2 × 10 18 atoms/cm 2 , with the gas uniformly distributed over the 4 cm length of the cell and less than 1% residual gas outside the cell. This design eliminated scattering from the end caps of the target cell, a problem endemic to previous measurements of the proton charge radius in electron scattering experiments, and permitted a precise, model-independent extraction of rp by reaching unprecedentedly low values of Q2 , the square of the electron’s transfer of four-momentum to the proton.- Hydrogen target
- Gas target
- Proton charge radius
References(16)