Test of a small prototype of the COMET cylindrical drift chamber
Jun 4, 2021
22 pages
Published in:
- Nucl.Instrum.Meth.A 1015 (2021) 165756
- Published: Nov 1, 2021
e-Print:
- 2106.02316 [physics.ins-det]
DOI:
- 10.1016/j.nima.2021.165756 (publication)
View in:
Citations per year
Abstract: (Elsevier B.V.)
The performance of a small prototype of a cylindrical drift chamber (CDC) used in the COMET Phase-I experiment was studied by using an electron beam. The prototype chamber was constructed with alternating all-stereo wire configuration and operated with the HeiC4H10(90/10) gas mixture without a magnetic field. The drift space–time relation, drift velocity, d/d resolution, hit efficiency, and spatial resolution as a function of distance from the wire were investigated. The average spatial resolution of 150 m with the hit efficiency of 99% was obtained at applied voltages higher than 1800V. We have demonstrated that the design and gas mixture of the prototype match the operation of the COMET CDC.Note:
- 22 pages, 14 figures, published in Nucl. Inst. Meth. A
- Drift chamber
- Helium-based gas
- Spatial resolution
- helium-based gas
- spatial resolution
- wire
- drift chamber
- helium
- carbon: hydrogen
- efficiency
References(24)