Measurement of reactor thermal neutron fluence of NTD-Ge by activation High-Purity Ge itself

Feb 24, 2023
e-Print:

Citations per year

0 Citations
Abstract: (arXiv)
In neutron transmutation doped germanium, the thermal neutron fluence of reactor irradiation is as high as 1018^{18}~n\cdotcm2^{-2}. For radiological safety reasons, general Co or Au neutron flux monitors cannot be easily used. We have experimentally demonstrated the feasibility of measuring the X-rays emitted by the NTD-Ge itself to determine the absolute thermal neutron fluence for the first time. A Micro-Megas Detector (MMD) and a Silicon Drift Detector (SDD) are set up to detect the tagging KX-rays with 9.2 keV and 10.3 keV cascading from the decays of 71^{71}Ge. Combined the detection efficiencies calculated by GEANT4, neutron fluence results given with proper accuracy by MMD and SDD are in a good agreement with each other.