Order-of-magnitude SNR improvement for high-field EPR spectrometers via 3D printed quasi-optical sample holders
May 2, 2023Published in:
- Sci.Adv. 9 (2023) 38, adi7412
- Published: Sep 22, 2023
e-Print:
- 2305.01820 [physics.ins-det]
View in:
Citations per year
Abstract: (American Association for the Advancement of Science)
Here, we present a rapidly prototyped, cost-efficient, and 3D printed quasi-optical sample holder for improving the signal-to-noise ratio (SNR) in modern, resonator-free, and high-field electron paramagnetic resonance (HFEPR) spectrometers. Such spectrometers typically operate in induction mode: The detected EPR (“cross-polar”) signal is polarized orthogonal to the incident (“co-polar”) radiation. The sample holder makes use of an adjustable sample positioner that allows for optimizing the sample position to maximize the 240-gigahertz magnetic field B1 and a rooftop mirror that allows for small rotations of the microwave polarization to maximize the cross-polar signal and minimize the co-polar background. When optimally tuned, the sample holder was able to improve co-polar isolation by ≳20 decibels, which is proven beneficial for maximizing the SNR in rapid-scan, pulsed, and continuous-wave EPR experiments. In rapid-scan mode, the improved SNR enabled the recording of entire EPR spectra of a narrow-line radical in millisecond time scales, which, in turn, enabled real-time monitoring of a sample’s evolving line shape.References(75)
Figures(11)