The Performance of thin NaI(Tl) scintillator plate for dark matter search
Dec, 2005
Citations per year
Abstract:
A thin (0.05cm) and wide area (5cmX5cm) NaI(Tl) scintillator was developed. The performance of the thin NaI(Tl) plate, energy resolution, single photoelectron energy and position sensitivity were tested. An excellent energy resolution of 20% (FWHM) at 60keV was obtained. The single photoelectron energy was calculated to be approximately 0.42 0.02keV. Position information in the 5cmx5cm area of the detector was also obtained by analyzing the ratio of the number of photons collected at opposite ends of the detector. The position resolution was obtained to be 1cm (FWHM) in the 5cmx5cm area.- thin NaI(Tl) scintillator
- energy resolution
- position sensitivity
- dark matter: search for
- WIMP nucleus: interaction
- scintillation counter: crystal
- sodium: iodine
- performance
- energy resolution
- spatial resolution
References(9)
Figures(0)