A 128-channel picoammeter system and its application on charged particle beam current distribution measurements

Sep 30, 2015
6 pages
Published in:
  • Rev.Sci.Instrum. 86 (2015) 11, 115102
  • Published: Nov 3, 2015
e-Print:

Citations per year

0 Citations
Abstract: (AIP)
A 128-channel picoammeter system is constructed based on instrumentation amplifiers. Taking advantage of a high electric potential and narrow bandwidth in DC energetic charged beam measurements, a current resolution better than 5 fA can be achieved. Two sets of 128-channel strip electrodes are implemented on printed circuit boards and are employed for ion and electron beamcurrent distribution measurements. Tests with 60 keV O3+ ions and 2 keV electrons show that it can provide exact boundaries when a positive charged particle beam current distribution is measured.
Note:
  • 6 piges, 8 figures; Submitted to Review of Scientific Instruments
  • Electrodes
  • Electric currents
  • Electrical resistivity
  • Resistors
  • Capacitors