Characterisation of Two PMT Models for the IceCube Upgrade mDOM
Aug 22, 20198 pages
Published in:
- PoS ICRC2019 (2020) 1022
Contribution to:
- , 1022
- ICRC 2019
e-Print:
- 1908.08446 [astro-ph.IM]
DOI:
Report number:
- PoS-ICRC2019-1022
Experiments:
View in:
Citations per year
Abstract: (SISSA)
The IceCube Upgrade will expand the IceCube Neutrino Observatory with nearly 800 new optical modules. A large fraction of these will be multi-PMT optical modules (mDOMs), featuring 24 PMTs pointing uniformly in all directions, providing an almost homogeneous angular coverage and providing an effective photosensitive area more than twice that of current IceCube optical modules. Two PMT models from different manufacturers are currently considered for use in the mDOM: a 3.5 inch PMT from HZC Photonics and a 3 inch PMT from Hamamatsu. Both PMTs have been characterized in terms of gain, timing, quantum efficiency and dark noise rate as a function of temperature. The obtained characterization results are presented here.Note:
- Presented at the 36th International Cosmic Ray Conference (ICRC 2019). See arXiv:1907.11699 for all IceCube contributions
- IceCube: upgrade
- efficiency: quantum
- photomultiplier
- optical
- observatory
- temperature
- noise
References(9)