Radon-implanted Po-214 and anomalous pulses in sodium iodide detectors for dark matter

2000
3 pages
Published in:
  • Phys.Lett.B 490 (2000) 6-8

Citations per year

20012006201120162020102
Abstract: (Elsevier)
214 Po is implanted into surfaces exposed to radon. The implantation rate could be sufficient to explain the rate of low-energy events with anomalously fast scintillation pulses that has been observed in sodium iodide crystals used in a search for WIMP dark matter. The rate should be closely coupled in time to the radon concentration.
  • 95.35.+d
  • 23.60.+e
  • 29.40.Mc
  • Radon
  • Radioactive background
  • Dark matter
  • dark matter
  • WIMP: search for
  • scintillation counter: crystal
  • scintillation counter: sodium-iodide