Citations per year

2010201320162019202102468
Abstract: (arXiv)
In the field of Double Beta Decay (DBD) searches, the use of high resolution detectors in which background can be actively discriminated is very appealing. Scintillating bolometers containing a Double Beta Decay emitter can largely fulfill this very interesting possibility. In this paper we present the latest results obtained with CdWO4 and CaMoO4 crystals. Moreover we report, for the first time, a very interesting feature of CaMoO4 bolometers: the possibility to discriminate beta-gamma events from those induced by alpha particles thanks to different thermal pulse shape.
  • 29.40.M
  • 23.40.B
  • 07.57.K
  • Double Beta Decay
  • Bolometers
  • Scintillating
  • CdWO4
  • CaMoO4
  • Pulse shape discrimination (PSD)
  • double-beta decay