Simulation Study of Energy Resolution with Changing Pixel Size for Radon Monitor Based on Topmetal-II{II}^- TPC

Dec 14, 2017
8 pages
Published in:
  • Nucl.Sci.Tech. 30 (2019) 1, 16
  • Published: Jan 2, 2019
e-Print:
DOI:

Citations per year

2020202120222023202410
Abstract: (Springer)
In this paper, we study how pixel size influences energy resolution for a proposed pixelated detector—a high sensitivity, low cost, and real-time radon monitor based on a Topmetal- II{II}^- time projection chamber (TPC). This monitor was designed to improve spatial resolution for detecting radon alpha particles using Topmetal- II{II}^- sensors assembled by a 0.35 μm CMOS integrated circuit process. Owing to concerns that small pixel size might have the side effect of worsening energy resolution due to lower signal-to-noise ratio, a Geant4-based simulation was used to investigate the dependence of energy resolution on pixel sizes ranging from 60 to 600 μm. A non-monotonic trend in this region shows the combined effect of pixel size and threshold on pixels, analyzed by introducing an empirical expression. Pixel noise contributes 50 keV full-width at half-maximum energy resolution for 400 μm pixel size at 1–4 σ\sigma threshold that is comparable to the energy resolution caused by energy fluctuations in the TPC ionization process ( 20\sim \,20 keV). The total energy resolution after combining both factors is estimated to be 54 keV for a pixel size of 400 μm at 1–4 σ\sigma threshold. The analysis presented in this paper would help choosing suitable pixel size for future pixelated detectors.
Note:
  • 8 pages, 7 figures
  • Geant4
  • Energy resolution
  • Pixel size
  • Radon monitor
  • Topmetal