The MR-TOF-MS isobar separator for the TITAN facility at TRIUMF

May 8, 2015
10 pages
Published in:
  • Hyperfine Interact. 235 (2015) 1-3, 97-106
Contribution to:
  • Published: 2017
    and
  • Published: May 8, 2015

Citations per year

2015201820212024202502468
Abstract: (Springer)
At TRIUMF’s Ion Trap for Atomic and Nuclear Science (TITAN) a multiple-reflection time-of-flight mass spectrometer (MR-TOF-MS) will extend TITAN’s capabilities and facilitate mass measurements and in-trap decay spectroscopy of exotic nuclei that so far have not been possible due to strong isobaric contamination. This MR-TOF-MS will also enable mass measurements of very short-lived nuclides (half-life > 1 ms) that are produced in very low quantities (a few detected ions overall). In order to allow the installation of an MR-TOF-MS in the restricted space on the platform, on which the TITAN facility is located, novel mass spectrometric methods have been developed. Transport, cooling and distribution of the ions inside the device is done using a buffer gas-filled RFQ-based ion beam switchyard. Mass selection is achieved using a dynamic retrapping technique after time-of-flight analysis in an electrostatic isochronous reflector system. Only due to the combination of these novel methods the realization of an MR-TOF-MS based isobar separator at TITAN has become possible. The device has been built, commissioned off-line and is currently under installation at TITAN.
  • Isobar separation
  • MR-TOF-MS
  • ISOL
  • Low-energy RIB
  • Mass measurements
  • Exotic nuclides