A Solenoidal electron spectrometer for a precision measurement of the neutron beta-asymmetry with ultracold neutrons

Jun, 2008
30 pages
Published in:
  • Nucl.Instrum.Meth.A 595 (2008) 587-598
e-Print:

Citations per year

2008201220162020202201234
Abstract: (arXiv)
We describe an electron spectrometer designed for a precision measurement of the neutron β\beta-asymmetry with spin-polarized ultracold neutrons. The spectrometer consists of a 1.0-Tesla solenoidal field with two identical multiwire proportional chamber and plastic scintillator electron detector packages situated within 0.6-Tesla field-expansion regions. Select results from performance studies of the spectrometer with calibration sources are reported.
  • 29.40.Cs
  • 29.30.Aj
  • 29.40.Mc
  • 23.20.En
  • Neutron beta -decay
  • Low-energy electron magnetic spectrometer
  • Ultracold neutrons
  • n: polarization
  • n: semileptonic decay
  • asymmetry